THE METER OF CHOICE
Since the late 1970’s, the Vortex direct steam flow meters have been acknowledged as the industry standard for direct steam flow measurement
General Specifications
- Meter willl consist of a full-bore body, vortex shedding bar and remote or integral electronics.
- Meter available with local or remote indication or blind housings.
- Meter available with pulse or analog (4-20 mA) outputs.
- Instantaneous and totalized flow available at local indicator or remotely through outputs.
- Meter measures flow using the Karman vortex shedding principle.
- Vortices induce pulses generated from two piezoelectric crystals hermetically sealed & imbedded in shedder bar.
- Meter electronics shall be capable of direct MASS flow computation for saturated steam without external inputs.
- Meter provides one button autotuning for setting noise immunity circuits.
- Operating pressure/temperature of meter shall be (Vacuum to 600 psig)/( -40°F to 500°F)